Recall Z-2202-2020 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Agfa N.V.

AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to

DateApr 6, 2020
StatusTerminated
ReasonUnder specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible
Root causeSoftware design
ActionAgfa issued Urgent Field Safety Notice via email on 4/21/20 to consignees, letter states reason for recall, health risk and action to take: Acknowledgment, via a feedback form or email, that the information was received and understood was requested from the consignees.
Quantity13 units
DistributionUS Nationwide distribution including in the sates of FL, IL, IN, LA, MA, MO, OH, SC, VA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180589DR 800 with MUSICA Dynamic
Agfa Healthcare N.V., Apr 5, 2018