Recall Z-2162-2020 ยท ELC Scaler, Ultrasonic

Sonendo Inc

GENTLEWAVE REF: FG-002-0001

DateOct 23, 2019
StatusTerminated
ReasonThe console would continue to run for extended period when the foot pedal was released.
Root causeDevice Design
ActionOn 10/23/2019, "Urgent: Medical Device Recall" of a foot pedal, the firm provided information on how to recognize the device malfunction as: If the treatment fluid is continuously cycling after the foot pedal is released, this is a sign that the foot pedal is not working properly. The firm instructed customers to replace the current foot pedal with the new foot pedal that was sent with this notification. and instructed to: Attach the new foot pedal to the Foot Pedal connection located at the lower rear of the Console as instructed per IFU. The firm instructed customers to return the current fo
Quantity460 foot pedals
DistributionU.S.: OR, CA, NY, SD, CO, AL, WA, IN, TX, NM, IL, NC, AZ, TN, KS, MT, FL, UT, OK, ID, OH, ME, WV, CT, GA, MO, MI, NE, KY, VT, ND, NJ, DC, LA, NY, MA, VA, MN, SC, MD, PA, NH, NV, NE, OR, MS, WI, AK, IA. No foreign consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190359Sonendo GentleWave System
Sonendo, Inc., Mar 16, 2019