Recall Z-2117-2020 ยท LCD Enzyme Immunoassay, Gentamicin

Abbott Gmbh & Co. KG

ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.

DateMay 12, 2020 (initiated Apr 24, 2020)
StatusOpen, Classified
ReasonAbbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.
Root causeUnder Investigation by firm
ActionOn April 24, 2020, Abbott issued a Product Correction letter to all customers who received shipments of the three impacted lots of ARCHITECT iGenatamicin Reagent kits. U.S. customers were sent the letter via FedEx priority overnight express. In the U.S., the customers will be requested to return to Abbott a Customer Reply Form acknowledging the receipt or a request for assistance.
Quantity6087 Reagent Kits
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MT, ND, NE, NJ, NV, NY, OH, OK, PA, RI, SC, TN,TX, UT, VA, VT, WA, WI, PR. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Ba
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102699ARCHITECT IGENTAMICIN REAGENTS; ARCHITECT IGENTAMICIN CALIBRATORS
Abbott Laboratories, Apr 1, 2011