Recall Z-2372-2020 ยท MDT Elisa, Trichinella Spiralis

BIOSENSE WEBSTER

PENTARAY NAV eco HIGH-DENSITY MAPPING CATHETER ELECTRODE EA 22 Rx only STERILE EO - Product Usage: The Biosense Webster PENTARAY NAV High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the

DateApr 22, 2020
StatusTerminated
ReasonMapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
Root causeError in labeling
ActionOn 4/22/2020 the firm issued an "URGENT Medical Device Correction" letter via UPS/certified mail, informing them that as a result of 3 complaints in 2019 in which a catheter became entrapped in the cardiac valvular apparatus while the user was attempting to create an electroanatomic map, leading to the need for surgical intervention for removal, the firm is performing a correction to the products. The Notification Letter informs customers: Following are learnings that users of this product should be aware of to avoid or address entrapment: " Repeated rotation of the catheter near the valvular
Quantity435590 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI IL, VA,, NE, OK, NJ, CT, KS, HI, CO, ND, RI, ID, MS, VT, DC and WV. The
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123837PENTARAY NAV ECO HIGH-DENSITY MAPPING CATHETER
Biosense Webster, Inc., Mar 5, 2013