Recall Z-2089-2020 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Howmedica Osteonics Corp.
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 58 Right, Sterile, Catalog # 504-02-58D-R, UDI (01)07613327012446(17)
| Date | May 7, 2020 (initiated Apr 16, 2020) |
| Status | Terminated |
| Reason | The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris. |
| Root cause | Packaging |
| Action | Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued |
| Quantity | 4895 |
| Distribution | Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New Hampshire, Illinois, New Jersey, Virginia, Ut |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K151264 | Restoration Anatomic Shell Stryker Orthopaedics, Aug 10, 2015 |