ClearanceBrief
Recall Z-2086-2020 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Howmedica Osteonics Corp.

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 56 Left, Sterile, Catalog # 504-02-56D-L, UDI (01)07613327012361(17)

DateMay 7, 2020 (initiated Apr 16, 2020)
StatusTerminated
ReasonThe Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Root causePackaging
ActionStryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued
Quantity4895
DistributionWorldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New Hampshire, Illinois, New Jersey, Virginia, Ut
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151264Restoration Anatomic Shell
Stryker Orthopaedics, Aug 10, 2015