Recall Z-2206-2020 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Inc.

Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001

DateJun 3, 2020 (initiated Apr 10, 2020)
StatusTerminated
ReasonInternal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.
Root causeProcess control
ActionMedtronic initiated verbal communication with affected consignees on April 10, 2020 via phone call and provided affected consignees with an "Urgent Field Safety Notice" via email between the dates of 04/24/20 - 4/30/20. . In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Immediately identify and quarantine all unused affected pumps in your inventory. 2. Return all unused affected pumps in your inventory to Medtronic. Your Medtronic Representative can assist you in the return and replacement of this pump as necessary. 3. For
Quantity24 Pumps
DistributionNo US Distribution. OUS: Netherlands, France, and Sweden
FDA recordOpen on fda.gov