ClearanceBrief
Recall Z-2435-2020 ยท BYE Attachment, Breathing, Positive End Expiratory Pressure

Westmed, Inc.

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.

DateJan 15, 2019
StatusTerminated
ReasonA repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
Root causeDevice Design
ActionThe firm sent a notification letter via email and it was issued on 1/15/2019. The following are actions given by the recalling firm to be taken by the customer/user: 1. If you are using the Neo-Vent Circuit with a 10 mm ID connection system (such as the F&P Neopuff System) the connection is compatible with all product lots. You may continue to use it with no issues. 2. If you are using the Neo-Vent Circuit with a 15 mm OD connection system (such as the GE Panda or Giraffe System), the connection may not be compatible with the list of LOT numbers. Quarantine the product and contact via
Quantity35 cases
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K923618PEEP VALVE
International Respiratory Systems, Inc., Aug 3, 1992