Recall Z-2128-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

DateMay 4, 2020 (initiated Apr 17, 2020)
StatusTerminated
ReasonThe unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
Root causeDevice Design
ActionURGENT: MEDICAL DEVICE CORRECTION notification letters dated 4/17/20 were sent to customers. What is the impact to the operation of the system and what are the possible risks? There is no impact to the operation of the system, however, unplugging the main cable can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector which can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. How was the subject identified and what is the root cause? The issue was identified by the manufacturer during system assembly when the plug of
Quantity20
DistributionUS Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181550Cios Spin
Siemens Medical Solutions USA, Inc., Oct 30, 2018