Recall Z-2287-2020 ยท HQL Intraocular Lens

Alcon Research LLC

Acrysof, IQ IOL w/UltraSert System, AcrySof IQ ASPHERIC IOL, 20.5 D, SP ACRYLIC FOLDABLE LENS, w/UltraSert DELIVERY SYSTEM, UV w/BLUE LIGHT FILTER. 13.0mm. LENGTH, 6.0mm ANTERIOR ASYMMETRIC BICONVEX OPTIC, PLANAR HAPTICS.

DateApr 17, 2020
StatusTerminated
ReasonIncorrect IOL diopter
Root causeProcess control
ActionOn 04/17/2020, the firm sent an "Urgent: Voluntary Medical Device Removal' letter to customers via overnight mail that the firm became aware that there was a potential problem relating to the IOL of wrong sized diopter, in a pre-loaded delivery system. Should a patient be implanted with an IOL having incorrect optical power, a refractive error would result. Patients with a refractive error may experience symptoms from mild blurred vision to decreased vision, which may require additional interventions such as spectacles or surgical intervention to correct. The firm is instructing customers
Quantity84 units
DistributionUS Nationwide distribution including in the states of TX, PA, CT, IA, MI, NE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.