Recall Z-2195-2020 ยท FJK Set, Tubing, Blood, With And Without Anti-Regurgitation Valve

Baxter Healthcare Corporation

Gambro Cartridge Low Volume Blood Transport System for Hemodialysis Blood Set - Product Usage: is intended for single use in a hemodialysis treatment using Dialysis Delivery Systems.

DateApr 17, 2020
StatusTerminated
ReasonPotential disconnection of tubing set.
Root causeProcess control
ActionBaxter Healthcare notified customers on about 04/17/2020 via "Urgent Device Correction" letter. Customers were instructed to do the following: 1. Prior to use, thoroughly inspect each connection of the set to check for any detachments in the tubing. Operators may continue to use affected sets if no detachments are observed. Additionally, per the Instructions For Use (IFU), users should observe carefully for leaks during priming and use, and examine the tubing carefully to be certain that all connections are secure, all lines are unobstructed and that there are no kinks or leaks in the tubi
Quantity19,800 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Countries of Australia, Austria, Bahrain, Bel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070414GAMBRO CARTRIDGE BLOOD SETS
Gambro Renal Products, Oct 12, 2007
K100364GAMBRO CARTRIDGE BLOOD SET LOW WEIGHT- LOW VOLUME, MODEL 003422-520
Gambro Renal Products, Inc., Nov 3, 2010