Recall Z-2129-2020 ยท BSP Needle, Conduction, Anesthetic (W/Wo Introducer)

TELEFLEX MEDICAL INC

Sprotte Needle, REF 32115130C w/ Intro 22 G x 3.5", distributed by Teleflex Medical The SPROTTE lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/discography procedures.

DateOct 11, 2019
StatusOpen, Classified
ReasonTeleflex received the attached Medical Device Recall Notice from Pajunk Medical. Teleflex Medical is a distributor of Pajunk products. According to the notice, Pajunk internally identified a packaging problem that may affect certain batches of the SPROTTE lumbar puncture cannulas with introducer. Due to this problem Pajunk cannot guarantee the sterility of the affected batches.
Root causeProcess control
ActionThe firm, Teleflex Medical, initiated the recall by letter on 10/11/2019. The notice requested that the consignees take actions as follows: Distributors: 1. Immediately discontinue distribution and quarantine any products with the product codes and lot numbers listed in the letter. 2. Using the provided customer letter and recall acknowledgement form templates, communicate this recall to any of your customers who have received product included within the scope of the recall. 3. Have the customers return any affected product to you, together with a completed Recall Acknowledgement Form, for co
Quantity4227 devices
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911260ATRAUMATIC STANDARD SPROTTE NEEDLE
Pajunk GmbH, Oct 21, 1991