Recall Z-2038-2020 ยท LXH Orthopedic Manual Surgical Instrument

Pega Medical Inc.

Orthopedic manual Surgical Instrument handle for Simple Locking Intra Medullary (SLIM) System - Product Usage: The SLIM handle is the grip for the SLIM driver in orthopedic surgical procedures.

DateApr 3, 2020
StatusTerminated
ReasonThe lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
Root causeNonconforming Material/Component
ActionOn 04/03/2020 Pega Medical Inc. sent out an email which included a Field Safety Notice letter to their US consignee identifying the problem, the actions to be taken by the customer/user in order to prevent risks for patients/users, and the actions planned by Pega Medical Inc. to correct the problem. PRODUCTS: The Handle of the Driver of the Simple Locking intramedullary System (SLIM) WITH CATALOG # SLM-HND100 which was distributed to the identified US consignee on 12/24/2019, with lot #8204-1 marked on the part (the driver handle) and with packaging labeled with lot# 8204-01. ACTION TO BE TAKE
Quantity3 units
DistributionUS Nationwide Distribution including in the states of Florida. No Foreign Consignees listed.
FDA recordOpen on fda.gov