Recall Z-2111-2020 ยท FMZ Incubator, Neonatal

GE Healthcare, LLC

Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.

DateMay 22, 2020 (initiated Apr 2, 2020)
StatusTerminated
ReasonDuring Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Root causeDevice Design
ActionGE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.
Quantity204 units.
DistributionUS Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.