ClearanceBrief
Recall Z-2185-2020 ยท MQV Filler, Bone Void, Calcium Compound

Osteomed, LLC

OSTEOVATIONEX 5CC IMPACTFORMULA. REF/UDI: 390-0205/00813845020047 - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.

DateMay 29, 2020 (initiated Apr 2, 2020)
StatusTerminated
ReasonIncorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
Root causeVendor change control
ActionOn 04/02/2020, an URGENT Medical Device Recall letter was sent to all consignees affected by the recall. OsteoMed has requested that affected customers: (1) review this letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or fax; and (2) return unused inventory via Federal Express (using OsteMed account)
Quantity300
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA. MA, MD, ME. MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY and the countries of AUSTRALIA, ITA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100986CALLOS BONE VOID FILLER, SKAFFOLD NEXT GENERATION BONE VOID FILLER, OSTEO VATION
Skeletal Kinetics, LLC, Jul 20, 2010