Recall Z-2361-2020 ยท JJG Analyzer, Chemistry, Centrifugal, For Clinical Use

Abaxis Inc

Piccolo Xpress chemistry analyzer, Model No. 1100-1001, Software version 2.1.55 (NGA)/3.1.35(Jen II)

DateDec 11, 2019
StatusTerminated
ReasonIncorrect reference ranges of analytes.
Root causeSoftware design
ActionOn January 9, 2020, the firm sent an "Urgent Medical Device Recall' to their customers, regarding the chemistry analyzer containing software versions software versions 2.1.55/3.1.3. 5. This software failure may cause the product to display the incorrect reference ranges for certain analytes when gender selection is not set on the device. Erroneous reference values may result in an incorrect diagnosis or treatment plan.The firm informed customers to immediately stop using the devices and return it to the firm. On January 21, 2020, the firm instructed their customers to check the enclosed list
Quantity15
DistributionAll U.S. Consignees, TX, PA, FL, OK, KY, CA, VA No OUS Consignees.