Recall Z-2046-2020 ยท JTS Stains, Microbiologic

Becton Dickinson & Co.

BD BBLTM Gram Crystal Violet 250mL-- IVD Primary stain for Staining Microorganisms, Catalog Number: 212525 - Product Usage: Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.

DateApr 14, 2020
StatusTerminated
ReasonMay exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
Root causeUnder Investigation by firm
ActionBD issued Urgent Medical Device Notification to Customers/Distributors on 4/14/20 via FedEX stating reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific Catalog and lot numbers listed above. Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form. Even if you do not have any o
Quantity10,568 bottles
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Taiwan, Hong Kong, Chile, EMEA, Thailand, China, Japan, Australia, New Zealand, Korea, Costa Rica, Mexico, Canada, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.