Recall Z-2338-2020 ยท LLZ System, Image Processing, Radiological

Philips Healthcare Informatics, Inc.

IntelliSpace PACS 4.4, Product number 837507 - Product Usage: intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians.

DateApr 19, 2019
StatusCompleted
ReasonImages may potentially become corrupt while using the system.
Root causeDevice Design
ActionThe recalling firm issued a customer information letter beginning April 29, 2019, via FedEx (for the U.S. and Canadian customers) informing the consignee of the issue, their site was affected, and a time needed to be scheduled for an onsite visit.
Quantity26 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of AZ, CA, FL, LA, MA, MI, MO, NC, NH, NY, OR, and TX. There was government distribution but no military distribution. The countries of Canada, Brazil, Ireland, New Zealand, United Arab Emirates, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111804INTELLISPACE PACS 4.X
Philips Healthcare Informatics, Inc., Jul 28, 2011