ClearanceBrief
Recall Z-2027-2020 ยท HWJ Awl

K2M, Inc

K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90167, UDI # 10888857108806, Size 28x8mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.

DateApr 24, 2020 (initiated Apr 13, 2020)
StatusTerminated
ReasonThe Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Root causeDevice Design
ActionOn April 13, 2020, a Product Recall Notification letter was sent to all affected customers (Distributors and Branches/Agencies). The firm instructed customers to review the notification and then, inform all sales representatives and operation teams within the Branch/Agency. Next, the customers were told to check their inventory to locate the product listed and complete the Business Reply Form, even if no affected products were identified. If product was found on hand, the customers were instructed to indicate the quantity of units on the form and to return the product to the firm.
DistributionWorldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.