Recall Z-1825-2020 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta Inc

Elekta MONACO RTP Sytem, radiation treatment planning software system - Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The

DateApr 16, 2020
StatusOpen, Classified
ReasonThe Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose.
Root causeSoftware design
ActionIn April of 2020, Elekta Inc. prepared an Urgent/Important Field Safety Notice identifying this issue, the actions that should be taken by the user/customer in order to prevent risks for patients and/or users, and the actions planned by Elekta to correct the problem. PRODUCT: The Monaco RTP System with versions 5.40 or 5.40.01 installed in or after June of 2018. ACTIONS TO BE TAKEN BY THE CUSTOMER OR USER: Customers/users are instructed to follow their standard clinical practice of reviewing plans, including the review of contours and volumes. DVHs that show unexpected overdose or under-dose
Quantity96 devices
DistributionWorldwide distribution - US Nationwide Distribution in the states of IA, NJ, NY, PA, TX, WI and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183037Monaco RTP System
Elekta, Inc., Dec 4, 2018