Recall Z-1806-2020 ยท KPR System, X-Ray, Stationary

Philips North America, LLC

Philips TraumaDiagnost- a Stationary X-Ray System (Bucky System) Model: 720013 - Product Usage: general radiography.

DateApr 7, 2020
StatusTerminated
ReasonIf the shaft breaks due to material fatigue, the trauma arm can fall down and may result in injury.
Root causeDevice Design
ActionPhilips issued Urgent Field Safety Notice (FCO72000015) sent 4/09/20 via certified mail states reason for recall health risk, action to take: The IfU (Instruction for Use) require the operator to not use the system and call for service, if a mechanical defect or malfunction is suspected. As soon as a deformation is determined, the system may no longer be used and call the service engineer. Should you feel uncertain regarding these instructions, please contact Philips. If you would like any further information or support concerning this issue, please contact your local Philips representative.
Quantity346 units (5 US and 341 OUS)
DistributionWorldwide distribution - US Nationwide distribution including the states of CA, ME, RI, TX, WI and the countries of Canada, Australia, Austria, France, Germany, Greece, Hungary, Ireland, Netherlands, Portugal, Slovenia, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945278PHILIPS BUCKY DIAGNOST FAMILY
Philips Medical Systems North America, Inc., Dec 14, 1994