ClearanceBrief
Recall Z-2198-2020 ยท CEM Electrode, Ion Specific, Potassium

Inpeco S.A.

FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.

DateApr 7, 2020
StatusOpen, Classified
ReasonPotential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
Root causeSoftware design
ActionUrgent: Medical Device Correction dated 03/30/2020 were sent to customers. The impacted modules are the Aliquoter Modules (Inpeco Part Number FLX-212) with a firmware version prior to the following: AQMb_3-3-0.H86; AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86; xAQMb_1-1-0.elf; Action to be taken by the user To avoid the risk of contamination take the following precaution: 1) Discard the Primary Tube flagged with Clot Detection error or manage it according to your laboratory guidelines considering that it may be diluted. 2) Call Service Assistant in case the frequency of the Clot Detection Error i
Quantity71 systems
DistributionUS Nationwide distribution including in the states of IL, NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121012FLEXLAB 3.6, ACCELERATOR A3600
Inpeco S.P.A., Aug 31, 2012