Recall Z-1712-2020 ยท KGI Densitometer, Bone

GE Healthcare, LLC

Lunar iDXA Pro/Lunar iDXA Advance (iDXA with enCORE v18) bone densitometer

DateApr 24, 2020 (initiated Dec 3, 2019)
StatusOpen, Classified
ReasonAn issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from
Quantity123 in total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191112GEHC DXA Bone Densitometers with enCORE version 18
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC, Sep 19, 2019