Recall Z-2004-2020 ยท CEM Electrode, Ion Specific, Potassium

Inpeco S.A.

Abbott Accelerator a3600, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: ACP-208-00 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

DateApr 7, 2020
StatusTerminated
ReasonWhen analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Root causeSoftware design
ActionOn April 7, 2020, the firm sent an Urgent Medical Device Correction letter to all impacted customers and instructed them to return the Customer Letter Receipt Confirmation and Implementation Check form within 15 days. The customers were asked to confirm that they had read, understood, and applied all the applicable recommended actions. The impacted Customers were told that they would be contacted by their service representative to schedule the system update.
Quantity273
DistributionWorldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121012FLEXLAB 3.6, ACCELERATOR A3600
Inpeco S.P.A., Aug 31, 2012