Recall Z-1803-2020 ยท IYE Accelerator, Linear, Medical

Viewray, Inc.

ViewRay System: Model No. 10000 and 20000 for radiation treatment. PN11440 MRIdian System Treatment Planning and Delivery System (TPDS) software

DateNov 18, 2019
StatusTerminated
ReasonRegistering dose and structures in the treatment delivery workflow could result in an alignment discrepancy between the imported previously delivered dose and the displayed patient anatomy and structures. The MRIdian TPDS software shows this misalignment to the user in the predicted dose-volume histogram (DVH). The misalignment impact may not be obvious to the user. During plan re-optimization the previously delivered dose would be incorrectly accounted for by the software. This may result in unexpectedly higher or lower dose than the intended dose calculated for the subsequent treatment plan
Root causeSoftware design
ActionOn 11/18/2019, the firm send an "Urgent Device Correction" letter to customers via email to inform them firm has become aware of an issue impacting the MRIdian System Treatment Planning and Delivery System (TPDS) software. The issue relates to an optional feature related to using previously delivered dose to make a new plan on MRIdian. The firm determined that in some cases dose previously delivered to a patient may not be accurately registered with a daily scan. The firm is instructing its customers to: 1) Do not: Sum previously delivered treatment plans in the treatment delivery workflow
Quantity32
DistributionWorldwide distribution. US states of MO, WI, CA, FL, MI, NY, MI, IL, and MA. South Korea, Netherlands, Italy, Japan, Germany, UAE, Turkey, France, Israel, Denmark, Switzerland, and United Kingdom. The distributors are assigned to applicable regions. Firm send the Urgent Device Corrections to t
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111862VIEWRAY SYSTEM
Viewray, Incorporated, May 22, 2012
K170751MRIdian Linac System with 138-leaf Collimator
Viewray, Incorporated, Jun 7, 2017