Recall Z-0567-2021 ยท GDC Table, Operating-Room, Electrical

Mizuho America, Inc.

Microsurgical Operating Room Tables;7300BX (18-351-70).

DateSep 11, 2019
StatusOpen, Classified
ReasonUnintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.
Root causeDevice Design
ActionThe firm's representatives notified their U.S. consignees regarding the recall notification by phone and a separate FedEx letter was sent (08/09/2019 - date the letter was issued). The Customer Notification Letter "IMPORTANT VOLUNTARY RECALL NOTIFICATION", informed customers of the two (2) types of errors: 1) Unintended Tilting of the Tabletop - Impurities in the hydraulic oil has caused minor leaking inside a control valve causing the table to tilt after remaining static for long periods of time. 2) Incorrect Operation due to Infrared Operating Signal Errors - The table can inadvertently mi
Quantity10
DistributionUS: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.