Recall Z-1765-2020 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the

DateApr 6, 2020
StatusTerminated
ReasonIt was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.
Root causeRadiation Control for Health and Safety Act
ActionOnly one user in the United States is affected by this corrective action. The Customer Safety Advisory Notice was sent to the affected customer. Any additional communication regarding this corrective action will include the following statement: The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included
Quantity399 (wordwide)
DistributionWorldwide Distribution