Recall Z-1934-2020 ยท KPR System, X-Ray, Stationary

Cuattro LLC

CuattroDR System Software, REF: SWR-00069-MD, UDI: B269SWR00069MD0

DateOct 15, 2019
StatusTerminated
ReasonThere is a potential that an image from a previous patient study to show up in a later patient study. This may result in misdiagnosis of the correct patient.
Root causeDevice Design
ActionOn October 15, 2019, The firm contacted all affected consignees via phone call. In addition to informing consignees about the recall. The firm provided the following information to consignees: . To prevent a reoccurrence of this problem Cuattro has updated the settings file in the program data folder on your acquisition workstation. This update to the file will prevent any backup images from being downloaded to the acquisition workstation. This was completed for your system on September 6, 2019, service ticket number 70002871
Quantity9 units
DistributionUS: NJ, MI, SC, NC, MI, AR, OUS: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161937CuattroDR
Cuattro, LLC, Oct 6, 2016