Recall Z-1889-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Inpeco S.A.

Siemens Aptio Automation Centrifuge Module, tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf

DateApr 7, 2020
StatusOpen, Classified
ReasonThe FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Root causeSoftware design (manufacturing process)
ActionThe firm sent an Urgent Medical Device Correction letter to all affected customers on March 30, 2020. The letter instructed the customers to complete and return the Customer Letter Receipt Confirmation and Implementation Check form. The firm explained that their regular service provider will contact them to schedule the firmware version upgrade.
Quantity313
DistributionWorldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan