ClearanceBrief
Recall Z-2132-2020 ยท HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer

Smith & Nephew, Inc.

JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUM

DateJan 28, 2020
StatusTerminated
ReasonThe spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.
Root causeProcess design
ActionThe firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.
Quantity200 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190085Journey II Unicompartmental Knee System
Smith & Nephew, Inc., Feb 11, 2019