Recall Z-2065-2020 ยท NGT Saline, Vascular Access Flush

Becton Dickinson & Company

BD Posiflush SF Saline Flush Syringe (US Distributed Catalog 306553), UDI 50382903065535

DateApr 16, 2020
StatusTerminated
ReasonThis product has been confirmed to exhibit holes in the packaging, which impacts package integrity and potentially compromises a sterile field.
Root causeUnder Investigation by firm
ActionOn April 16, 2020, the firm initiated customer notification via Urgent Medical Device Recall letters. Customers were informed that the product had been confirmed to exhibit holes in the packaging, potentially compromising a sterile field. Customers were asked to do the following: 1. Immediately review your inventory for the catalog and lot numbers specified in this notification and discard all product subject to the recall. 2. Share this recall notification with any users of the product, within your facilities, to ensure they are also aware of this recall. 3. Complete and return the attached
Quantity16,099,310 units
DistributionNationwide domestic distribution. Foreign distribution to Austria, Belgium, Denmark, Finland, France, Germany, Italy, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, UK, Australia, Canada, Mexico, Malaysia, New Zealand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1534810.9% Sodium Chloride Injection, USP BD PosiFlush SF Syringe
Becton, Dickinson and Company, Apr 4, 2016