Recall Z-2591-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Shimadzu Medical Systems

Trinias DAP Meter, a component of the Digital Anglo System Trinias contained within the DAR-9500f - Product Usage: It is intended to be used for interventional procedures, such as invasive cardiology, interventional radiology, and interventional neuroradiology.

DateNov 4, 2019
StatusCompleted
ReasonThe DAP meters were found to be outside of tolerance.
Root causeDevice Design
ActionThe firm issued a Field Safety Notice dated 11/5/2019 via email on 11/4/2019 describing the issue in "Digital Angiography System Trinias." The local representative was going to correct the device at the consignee locations.
Quantity3 devices
DistributionUS Nationwide distribution including in the states of FL, KY, and LA. There was no government/military/foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123508TRINIAS
Shimadzu Corporation, Feb 26, 2014