Recall Z-1809-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America, LLC

Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

DateApr 29, 2020 (initiated Apr 3, 2020)
StatusTerminated
ReasonWhen a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.
Root causeSoftware design
ActionPhilips Healthcare issued Urgent Medical Device Correction letter via Field Action 2020-IGTBST-003 issued on April 3, 2020. Letter states reason for recall, health risk an action to take: To avoid this hazardous situation, the users should: BY CUSTOMER/ USER " Use XperGuide live guidance with the L-arm positioned in the XperCT scan position. " Move the L-arm back to the initial XperCT scan position if the system displays the warning Live 3D roadmap not possible. Please move stand to the : [acquisition position] position. At any stage during needle guidance, the user can perform a verification
Quantity22 units (6 US and 16 OUS); Expanded 3units; Total: 25
DistributionWorldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018