ClearanceBrief
Recall Z-2057-2020 ยท DSQ Ventricular (Assist) Bypass

Heartware, Inc.

Heartware HVAD Pump Implant Kit, Product (REF) Number 1104JP

DateApr 3, 2020
StatusOpen, Classified
ReasonThe Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Root causeDevice Design
ActionThe firm disseminated a Medical Device Safety Alert by letter beginning on 04/03/2020. The notice explained he potential for damage that can occur during the procedure. Specifically, the Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump. The letter also provided additional steps for assembly and attachment to reduce the risk of damage and tearing during the assembly procedure which was provided in Appendix A.
Quantity85 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.