Recall Z-2496-2020 ยท LLZ System, Image Processing, Radiological

Tomtec Imaging Systems Gmbh

TOMTEC-ARENA TTA2;software version no. 2.20 and lower; Picture archiving and communications system. Software Versions and applicable UDI: TTA 1.2 release prior to UDI requirement TTA2.00.01 release prior to UDI requirement TTA2.11.01 release prior to UDI requirement TTA2.11.02 release prior to UDI requirement TTA2.20.01 UDI: (01)04260361440143(11)160429(10)20.01 TTA2.20.10 UDI:(01)04260361440143

DateSep 23, 2019
StatusOpen, Classified
ReasonThe firm discovered a software issue associated with the Image-Com module/clinical application package (CAP) interface for 3D application in the ARENA software, where measurements from one study might be stored to another study.
Root causeSoftware design
ActionCustomers were sent the Urgent Medical Device Correction, Field Safety Notice letters, issued on 09/23/2019 via certified mail (USPS) with return receipt. The firm asked customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction Letter. The instructions describe how to delete all measurements if a Clinical Application Package (CAP) has been initiated and the clip/image is rejected. The instructions to the customers to avoid the software issue are as follows: If a CAP has been initiated and the clip/image is rejected and one of the messages de
Quantity62 units
DistributionUS:AL, AZ, CA, CT, IL, MA, MI,NC, NJ, NY, OH, PA, RI, TX, VA. OUS: Australia, Canada, Czech Republic, France, Germany, Italy, Latvia, Sweden, Switzerland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150122TOMTEC-ARENA TTA2
Tomtec Imaging Systems GmbH, Feb 13, 2015