ClearanceBrief
Recall Z-2138-2020 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer Medical ApS

ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

DateNov 21, 2019
StatusTerminated
ReasonThe analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
Root causeNonconforming Material/Component
ActionTwo Urgent Medical Device Recall Letters were sent to customers at the user level via email followed by a hard copy letter sent FedEx 2nd day with tracking and proof of delivery. The first letter, dated 11-21-2019, was intended for customers not utilizing the analyzer for measuring blood samples from newborn babies of up to four weeks of age. The letter informed customers of biased results for tbil, that the bias increases with tHb concentration, and that the increase is not linear. Customers were informed of the potential risks to patients, and were requested to complete the Recall Response F
Quantity544
DistributionWorldwide distribution - US Nationwide. Also distributed OUS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160153ABL90 FLEX PLUS
Radiometer Medical Aps, Nov 4, 2016