Recall Z-1795-2020 ยท GCB Needle, Catheter

Becton Dickinson & Company

CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number PIG1280T, UDI Number 01)10885403108402(17) 210930(10)0001304986

DateMar 26, 2020
StatusTerminated
ReasonThe 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle.
Root causeMixed-up of materials/components
ActionOn March 26, 2020, an Urgent Medical Device Recall Notification was sent to consignees stating that Lot #0001304986 of the Safe-T-Centesis 8 Fr Catheter Drainage Trays shipped between June 27, 2019 and September 6, 2019 should be discarded. In addition, they were instructed to complete the Customer Response form regardless of whether or not they have any of the product. Customers with discarded product were instructed to indicate the number of affected products that were discarded on the response form and replacements would be issued to those customers.
Quantity900
DistributionDomestic distribution to the following states: Alabama, Colorado, Illinois, Massachusetts, Missouri, Oklahoma, Rhode Island, Virginia, and Wisconsin.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.