Recall Z-1929-2020 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Encore Medical, LP

EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis

DateOct 22, 2019
StatusOpen, Classified
ReasonThe firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.
Root causeDevice Design
ActionThe firm requested its agents to identify and return affected devices.
Quantity86 units
DistributionUS Nationwide distribution including in the states of MN, VA, IL, IN, SC, N, TX, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190057DJO Acetabular System
Encore Medical L.P., Jun 7, 2019