Recall Z-2425-2020 ยท DZE Implant, Endosseous, Root-Form

Zest Anchors LLC

LOCATOR OVERDENTURE IMPLANT SYSTEM STERILE R Rx Only QTY 1 .

DateNov 22, 2019
StatusOpen, Classified
ReasonThe abutment provided with the Implant System may be out of specification, which could lead to a small gap occurring between the implant and abutment on assembly in the patient.
Root causeProcess control
ActionThe firm initially contacted their customers by phone between 11/20-22/2019 and then sent out Customer Notification Letters on 11/22/2019 via email. The Customer Notification Letter is informing customers: The firm has determined that the abutment provided with the Implant System may be out of specification, which could lead to a small gap occurring between the implant and abutment on assembly in the patient. The potential gap between the implant and the abutment could result in a space for bacteria to potentially harbor and grow, which could lead to an increase in future infection risk. T
Quantity154 kits
DistributionUS Nationwide Distribution in the state of: WI; VA; OH; IL; GA; NC; KS; NJ; CA; WA; OK; MO and Puerto Rico OUS: Switzerland, Germany, Mexico, Poland, and Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120198LOCATOR OVERDENTURE IMPLANT SYSTEM
Zest Anchors, LLC, Jun 6, 2012