Recall Z-1798-2020 ยท GEX Powered Laser Surgical Instrument

Boston Scientific Corporation

MoXy Laser Fiber, Material Number 0010-2400 - Product Usage: The fiber can be used for the surgical incision/excision, vaporization, ablation, hemostasis and coagulation of soft tissue.

DateMar 30, 2020
StatusTerminated
ReasonTo reinforce existing instructions within the Directions for Use (DFU) and provide further guidance which will be added to the DFU that may reduce the likelihood of metal cap and fiber tip break temperature related complaints.
Root causeUse error
ActionAn Urgent Medical Device Product Advisory - Immediate Action Required dated 3/30/20 was sent to customers. Existing MoXy" laser fiber Direction for Use to be Reinforced 1- Ensure the distance from the fiber to the tissue is approximately 2mm (1 to 3 mm). Reference: Table 2. Troubleshooting and Procedural Considerations in product DFU. 2- Do not bury the tip of the fiber into the tissue. Reduced irrigation flow may result in damage to the fiber. Reference: Caution in page 3 of product DFU. 3- Failure to position the saline solution at least 106cm or not opening the flow valve will decrease the
Quantity146,633
DistributionWorldwide distribution - US Nationwide and Global distribution.