Recall Z-2163-2020 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Hologic, Inc

Panther Fusion Extraction Reagent-S, Ref PRD-04331, IVD, CE, Rx Only consisting of FCR-S Fusion Capture Reagent - S, Nucleic Acid in a buffered solution and FER-S Fusion Enhancer Reagent-S (Lithium Hydroxide Solution), UDI: 15420045509214

DateOct 10, 2019
StatusOpen, Classified
ReasonReagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to invalid runs. This could result in incorrect results being reported and a delay in treatment.
Root causeNonconforming Material/Component
ActionCustomers who received product/affected lot were to be contacted via Urgent: Medical Device Recall Notification Letter, dated October 10, 2019 via FedEx overnight. In addition to notifying customers about the recall, the recall notification asked customers to take the following actions: 1) Immediately discontinue use of the lot of Panther Fusion Extraction Reagent-S kits and respective components listed in Table 1. 2) Immediately check inventory and segregate the Panther Fusion Extraction Reagent-S kit lot listed in Table 1. 3) Notified customers of the numbers of kits shipped to cust
Quantity47 kits
DistributionUS: AR, CA, FL, ID, IL, IN, LA, NM, NY, PA, TX, UT, VI, WA, WI OUS: AU, CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171963Panther Fusion Flu A/B/RSV Assay, Panther Fusion Universal Fluids Kit, Panther Fusion Assay Fluids I-S Kit, Panther Fusi
Hologic, Inc., Sep 26, 2017