Recall Z-1747-2020 ยท DQO Catheter, Intravascular, Diagnostic

Vascular Solutions, Inc.

Langston Dual Lumen Catheter 6F Model 5540. For delivery of contrast medium in angiographic studies

DateApr 27, 2020 (initiated Mar 16, 2020)
StatusTerminated
ReasonThe inner lumen of the Langston catheter may separate from the hub during pressure injection.
Root causeProcess control
ActionThe firm notified consignees that received affected product on March 26, 2020 via UPS Mail (US), email or mail (OUS). The firm s immediate correction is to recall the affected devices for destruction. The notice includes instructions on subrecall for any consignees that further distributed affected product.
Quantity30,877 units
DistributionWorldwide distribution. US nationwide, Austria, Australia, Belgium, Canada, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Mexico, Netherlands, Slovakia, South Africa, Spain, Sweden, Switzerland, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051395LANGSTON DUAL LUMEN PRESSURE MONITORING CATHETER
Vascular Solutions, Inc., Jun 24, 2005