ClearanceBrief
Recall Z-1679-2020 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus #5857912 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13

DateFeb 18, 2020
StatusTerminated
ReasonPotential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
Root causeUse error
ActionA Customer Safety Advisory Notice was sent to all affected consignees on 2/18/2020 via Adobe Mega Sign or USPS certified mail. The notice describes the interlock scenarios, provides user instructions, and warns the user of potential risks. The interlock scenarios are as follows: MONITOR 2: Monitor 2 total dose coincidence. Intent: This interlock is intended to prevent overdose when the primary dose monitory system fails to complete the treatment. TREATMENT TIME: Preset treatment time coincidence with actual treatment time. Intent: This interlock is intended to prevent overdose if preset trea
Quantity1012
DistributionWorldwide distribution. US consignees in the following states: Alabama, California, Delaware, Florida, Georgia, Iowa, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pen
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K882729MEVATRON, M 2 CLASS
Siemens Medical Laboratories, Inc., Sep 19, 1988
K982513MX MEVATRON
Siemens Medical Solutions USA, Inc., Jul 31, 1998