Recall Z-1960-2020 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci Xi Surgical System

DateNov 21, 2018
StatusTerminated
ReasonDue to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error 23087) and loss of arm functionality prior or during surgery.
Root causeProcess control
ActionOn 11/21/2018, the firm sent an "Urgent Medical Device Correction" Notification to its customers informing them that the firm had become aware of specific surgical systems were shipped with a manufacturing variation in the arm, which may result in a persistent recoverable fault (error 23087) with could occur prior or during a surgical procedure. The firm emphasized that as indicated in the User Manual, if this fault is persistent, ten the affected arm may be disabled. The firm is instructing that: 1. Inform surgeons and patient side assistants using the surgical system to read and understand
Quantity74 USMs containing affected rotors.
DistributionU.S.: CA, MD, NY, VA, IL, MI, DE, TX, FL, NJ, MA, KY and OH. O.U.S.: Australia, France, Germany and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall