Recall Z-1817-2020 ยท DPS Electrocardiograph
WELCH ALLYN, INC/MORTARA
Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.
| Date | Apr 29, 2020 (initiated Mar 18, 2020) |
| Status | Terminated |
| Reason | Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment. |
| Root cause | Device Design |
| Action | Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed. |
| Quantity | 78,521 units total |
| Distribution | US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K152944 | Xscribe Stress Exercise Testing Systems, Q-Stress Stress Exercise Testing Systems Mortara Instrument, Inc., Jan 6, 2016 |