Recall Z-1810-2020 ยท DPS Electrocardiograph

WELCH ALLYN, INC/MORTARA

Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.

DateApr 29, 2020 (initiated Mar 18, 2020)
StatusTerminated
ReasonImpacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Root causeDevice Design
ActionWelch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed.
Quantity78,521 units total
DistributionUS nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092663ELI 350 ELECTROCARDIOGRAPH
Mortara Instrument, Inc., Dec 14, 2009
K100127ELI 230 ELECTROCARDIOGRAPH
Mortara Instrument, Inc., Apr 16, 2010
K101115ELI 150 ELECTROCARDIOGRAPH MODEL ELI 150
Mortara Instrument, Inc., Jun 18, 2010
K101403ELI 250 ELECTROCARDIOGRAPH
Mortara Instrument, Inc., Jul 2, 2010
K102206ELI 10 ELECTROCARDIOGRAPH MODEL ELI 10
Mortara Instrument, Inc., Feb 18, 2011