Recall Z-1708-2020 ยท KPS System, Tomography, Computed, Emission

Shanghai United Imaging Healthcare Co., Ltd.

Positron Emission Tomograpy and Computed Tomography System, Model: uMI 550, with software version R001.3.0.0.750505, Rx Only, CE 0197, UDI: (01)06971576832026 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and dis

DateAug 26, 2019
StatusTerminated
ReasonPotential sporadic software bugs in R001.3.0.0.750505 software version may cause an issue with the PET acquisition raw data causing reconstruction problems during PET/CT scanning and could possibly cause rescans of patients with additional dose.
Root causeSoftware design
ActionOn August 26, 2019, the firm emailed consignee an "Customer Notification" recall letter. In addition to informing consignees about the recall letter, the notification ask consignees to take the following actions: 1. If the user finds that PET reconstruction failing, the PET should be rescanned immediately to complete a new reconstruction. 2. If this occurs, please contact the UIH Solutions (UIHS) customer Contact Center at telephone numbers (855) 221-1552 (toll free) or (832) 699-2799 (direct) to report the issue.
Quantity2 units
DistributionUS Nationwide distribution in the state TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182237uMI 550 PET/CT System
Shanghai United Imaging Healthcare Co., Ltd., Oct 12, 2018