Recall Z-1760-2020 ยท CEM Electrode, Ion Specific, Potassium

Radiometer Medical ApS

ABL90 FLEX Analyzer, Model Numbers 393-090, with software versions below 3.4MR2 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.

DateFeb 3, 2020
StatusTerminated
ReasonThe firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
Root causeSoftware design
ActionOn 2/3/2020, the firm provided affected consignees with a "Urgent: Medical Device Recall" letter via Email, and hard copy letter via FedEx 2nd day. In addition to providing the recall notification, consignees were asked to take the following actions: Check if your institution is using barcode types without a check digit, including e.g. patient ID or accession number, to be read on the ABL90. 1. If you use barcode types without a check digit, Radiometer recommends to either: - Enable the check digit on the barcode type currently used, or - Change type of barcode to one that includes a check
Quantity12,822 units
DistributionWorldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092686ABL90 FLEX SERIES
Radiometer Medical Aps, Aug 6, 2010