ClearanceBrief
Recall Z-1787-2020 ยท EZL Catheter, Retention Type, Balloon

Degania Medical Devices Pvt. Ltd.

Temperature Sensing 100% Silicone Foley Catheter, Part Number 102203101463DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.

DateMar 17, 2020
StatusTerminated
ReasonInadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Root causeNonconforming Material/Component
ActionAn Urgent Field Safety Notice dated 3/17/20 was sent to customers. Please immediately follow the Required Customer Actions set fo rth below. 1. Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Please promptly return the completed Recall Response fo rm to Degania Silicone by e-mail: kkayam@qco.net.
Quantity9137
DistributionWorldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063442ALL SILICONE FOLEY CATHETER, 2-WAY, 3-WAY, WITH TEMPERATURE SENSOR
Degania Silicone , Ltd., Mar 26, 2007