Recall Z-2131-2020 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

DateAug 19, 2019
StatusOpen, Classified
ReasonSome cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.
Root causeProcess control
ActionUrgent Medical Device Recall Letters (Field Safety Notices) were hand-delivered to U.S. customers on August 19, 2019 and OUS customers on August 20, 2019. Customers were asked to locate affected camera(s) in their inventory and inspect per the instructions attached to the Field Safety Notice throughout the life of the camera(s). If residual soil is present, customers were to return the affected camera(s) to the manufacturer using the standard Return Materials Authorization (RMA) process. Additionally, customers were requested to: iInform necessary hospital personnel about this issue; complet
Quantity305 units
DistributionDistribution US nationwide and South Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173906da Vinci SP Surgical System, EndoWrist SP Instruments, and Accessories
Intuitive Surgical, Inc., May 31, 2018